Infant Incubator For Neonatal Incubator ME-3000B

The infant incubator has a temperature control system and a humidity control system. The temperature control system includes two control working modes: air mode and skin temperature mode. Control of temperature heating output under air mode: The air temperature detected by the temperature sensor in the infant incubator is from the thermal output of the control heater; the control of the temperature heating output under the skin temperature mode: The skin temperature detected according to the temperature sensor comes from the heat output of the control heater. Humidity control system: The humidity detected by the humidity sensor in the infant incubator comes from the humidifier in the water tank.

Request Quote
Description

How To Use The Infant Incubator

First Point: Preparation
1. Do a good job of cleaning the whole machine and disinfection.
2. Lock the foot wheel of the whole machine to prevent the whole machine from moving.
3. The power cord -the end is inserted from the power supply input jack from the side of the chassis, and the other end is correctly inserted into the 220V/50Hz power socket, which is reliable.
4. Add an appropriate amount of distilled water to the water cup.
Second Point: Operation method
1. Turn on the power switch → the temperature controller enters the self -inspection status. Except for the ultra -temperature alarm indicator and the power -off alarm indicator, all indicators and digital tubes are on → All lights are destroyed → Display the temperature of the box and set the temperature value → If you do not press any keys, automatically enter the box temperature control status → The temperature value of the boot setting is to keep the setting value of the last shutdown.
2. Press the settings → Enter the box temperature setting status → set the temperature display window flickering. According to clinical needs, the temperature setting temperature is changed, push the add and subtract key to make the setting temperature display window consistent with the required value → Exit the setting state → enter the box temperature control state (if you do not press the box temperature key, wait for about 15 seconds to automatically enter the box temperature control status.
Third Point: Nursing method

When your baby cares, open the front panel lock, put down the front door panel, you can put in the baby, close the front door panel and lock the lock. When the baby needs to care, you can use the elbow to press the door lock and reach into the elliptical door for operation. In order to avoid the baby's cold, reduce the door opening time as much as possible. Hurry up and close the door after the completion.

Product feature
Air temp. and skin temp. servo-controlled by computer

Humidity servo-control by computer
>37ºC temperature setting
Various self-check alarms
Free-step mattress tilting adjustment
Removable humidity reservoir, easy to clean

With the transfusion shelf and tray
Added with guardrail
RS-232 connectorPhototherapy(option)


Infant Care Incubator ECOR005 parameters
Packing size(Up) 108*64*92cm
Packing size(Down) 108*64*84cm
N.W.(Up) 65Kg/51Kg
N.W.(Down) 55Kg/44Kg
Mattress temp. uniformity ≤0.8ºC
Environment relative humidity ≤80%
Atmospheric Pressure 500~1060hpa
Humidity control range(option) 30%RH-90%RH
Air control accuracy 0.5ºC