The infant incubator has a temperature control system and a humidity control system. The temperature control system includes two control working modes: air mode and skin temperature mode. Control of temperature heating output under air mode: The air temperature detected by the temperature sensor in the infant incubator is from the thermal output of the control heater; the control of the temperature heating output under the skin temperature mode: The skin temperature detected according to the temperature sensor comes from the heat output of the control heater. Humidity control system: The humidity detected by the humidity sensor in the infant incubator comes from the humidifier in the water tank.
Request Quote
Double walls hood, double door
Medical Infant Incubator Specifications | |
Electrical requirement | ~ 220V 50Hz or ~ 230V 60Hz |
Power consumption | ≤ 450VA |
Air mode temp. range | 25.0ºC~ 37.0ºC |
Skin mode temp. range | 34.0ºC~ 37.0ºC |
Air over temp. range | < 38.0ºC |
Air mode high temp. range | 37.1ºC~ 39.0ºC |
Skin mode high temp. range | 37.1ºC~ 38.0ºC |
High temp. mode alarm | < 40.0ºC |
Air mode alarm | ±3.0ºC |
Skin mode alarm | ±1.0ºC |
Skin sensor accuracy | ≤ ±0.3ºC |
Air control accuracy | ≤ 0.5ºC |
Mattress temp. uniformity | ≤ 0.8ºC |
Noise | ≤ 55dB(A) |